CT/ MRI Scanning Parameters:

CT/MRI scans that are taken for the purpose of 3D Printing should have following scan parameters or the closest approximation possible

  • Image Reconstruction Matrix: 512 x 512 or higher (Isotropic)
  • Slice thickness: 0.3 mm to 1.0 mm (As minimum as possible)
  • Feed per Rotation: 1.0 mm
  • Reconstructed Slice Increment: Less than or equal to slice thickness
  • Reconstruction Algorithm: Standard
  • Gantry Tilt: 0o
If you want to upload the DICOM data online:
  • Select all the DICOM files
  • Compress the files to .rar or .zip file format.
  • Upload the files on the company website www.anatomiz3d.com by following the instructions